KIMO MP120 Micromanometer User Manual

The KIMO MP120 is an intelligent micromanometer for measuring differential pressure ±0- 1000Pa, as well as air velocity with a Pitot tube (ordered separately).


KIMO MP120 Micromanometer User Manual - cover page
Brand
KIMO
Category
Test Set
Document type
User Manual
Language
DA
Pages
8
File format
PDF
File size
485 KB
Published
10 November, 2023
Updated
10 November, 2023
Model
MP120
MD5 checksum
76B5B4C3EF1E7DDB747C9EFBC3BFCC53

About this manual

The KIMO MP120 is a micromanometer for measuring differential pressure from ±0 to 1000 Pa and air velocity when a Pitot tube or other differential-pressure element is connected. It has a backlit display showing two values at the same time and supports the pressure and air velocity units listed in the document.

The manual explains starting and zeroing the instrument, holding and viewing minimum and maximum measurements, and configuring air velocity, units, auto-off time, key beep, coefficient, temperature, atmospheric pressure, and backlight. It also covers battery replacement with four AAA batteries.

What's inside

  • General description
  • Technical data
  • Instrument overview
  • Measurements with KIMO MP120
  • MP120 setup
  • Select pressure unit
  • Adjust auto-off time
  • Adjust coefficient
  • Adjust temperature
  • Battery replacement

Specifications

Pressure measuring range±0 to 1000 Pa / 100 mmH2O
Pressure accuracy±0.5% of RDG ±2 Pa
Resolution1 Pa
Air velocity measuring range2 to 40 m/s
Air velocity accuracy (2–5 m/s)±0.7 m/s
Air velocity accuracy (5–40 m/s)±0.5% of RDG ±0.3 m/s
Display4-line LCD, 50 x 36 mm
Battery4 x AAA, LR03 1.5 V, included
Dimensions148 x 71.5 x 35 mm
Weight220 g

Frequently asked questions

What pressure range does the KIMO MP120 measure?

It measures differential pressure from ±0 to 1000 Pa, equivalent to 100 mmH2O.

How do I perform a manual auto-zero adjustment?

Remove all pressure hoses, hold the instrument in free air, and press OK; Auto zero appears in the display.

How do I hold or freeze a measurement?

During measurement, press the HOLD-min/max button; HOLD appears and the measured result is frozen on the display.

What batteries are required?

The KIMO MP120 uses four LR03 1.5 V AAA batteries.