Technical specifications for KIMO differential pressure sensors and transmitters including output options, power supply, accuracy, and accessories.
This document provides technical specifications for KIMO differential pressure sensors and transmitters, covering models in Class 110, Class 210, and the PST series. It details output options such as 0-10 V or 4-20 mA, active and passive supply requirements, and available measuring units including Pa, mmH2O, inWG, and mBar.
| Output Options | 0-10 V or 4-20 mA |
|---|---|
| Active Supply (Class 110) | 24 Vac/Vdc (3-4 wires) |
| Passive Supply (Class 110) | 16 to 30 Vdc (2 wires) |
| Case Material | ABS VO IP65 |
| Accuracy (CP111) | ±1% of reading ±2 Pa |
| Accuracy (CP211/212) | ±0.5% of reading ±2 Pa |
| Temperature Input (Class 210) | Pt100 on terminal |
Available units include Pa, mmH2O, inWG, mmHG, daPa, kPa, hPa, and mbar.
Yes, the CP111 supports manual auto-zero by push-button or automatic auto-zero by solenoid valve.
The PST switch requires a supply of 24 Vac / Vdc.