The SI-1700 Auto/Manual Signal Generator is a versatile signal generator that can accept mA, 2 Wire, T/C, Volt, RTD or PMC signals. It uses a high-performance, high-precision 16-bit A/D converter to improve reliability. It can both manually adjust the output or receive the input value and output the isolated output.
This two-page leaflet describes the SI-1700 auto/manual generator, a panel-mount instrument that accepts multiple input signals and produces an isolated DC output. The model code is built as SI-1700 followed by digits for the input signal, output signal and power supply voltage; input types listed include thermocouples B, R, S, K, E, J, T and N, DPt100 and JPt100 RTDs, mV, Volt, mA, 2-wire 24V/4~20mA and PMC.
The document lists the main features, such as the 16-bit A/D converter, arbitrary output adjustment without an input, isolated output with output scaling, and the main functions Display Scaling, Sensor correction, Function (Linear, Root, Limit) and Output Scaling. It also gives the general specifications, a terminal connection diagram and the outline dimensions, and refers to page A-29 for the dimensional drawing. This manual is written in Korean.
| Input resistance | mA Type: 250Ω, Volt Type: 400kΩ, other types: 1MΩ |
|---|---|
| Sensor power | DC 24V / 50mA, within ±0.5% |
| Response speed | 100ms (Volt, mA), 200ms (T/C, RTD) |
| Filter function | NONE (FAST), 4, 8, 16, 32 |
| Allowable tolerance | 0.2% of Span at 23℃ |
| Linearity | 0.1% of Span |
| Output load resistance | 4~20mA DC (0~600Ω); 0~10V DC (1kΩ or more) |
| Insulation resistance | 100MΩ or more between input and output (1000V DC) |
| Withstand voltage | AC 1500V/min between input/output/power |
| Supply power | AC 85~264V (50/60Hz), 5VA or less; DC 24V ±10%, 200mA, 5VA or less |
| Weight | Approx. 250g (AC/DC) |
| Material / color | Flame-resistant resin / Gray |
| Outline dimensions | W 96 x H 48 x D 112 (mm) |
| Cutting size | W 92.5 x H 44.5 (mm) |
It is a multi-input unit that can receive mA, 2-wire, T/C, Volt, RTD or PMC signals.
The standard supply is AC 85~264V (50/60Hz) drawing 5VA or less; DC 24V is an option on the model code.
The response speed is 100ms for Volt and mA inputs and 200ms for T/C and RTD inputs.
For mV, Volt and mA types it offers Linear (passes the input unchanged), Root (square root for orifice flow measurement) and Limit (keeps the display at zero when it would go below zero).