MH-Z19B carbon dioxide gas sensor is a high-performance sensor with good selectivity and no oxygen dependence, long life and other features; Built-in temperature compensation; It also has serial, analog and PWM output, which is convenient to use.
The MH-Z19B is a general-purpose intelligent small sensor that uses the non-dispersive infrared (NDIR) principle to detect CO2 in air. It has good selectivity, no oxygen dependence, long life, built-in temperature compensation, and UART, analog (DAC), and PWM outputs. Its gold-plated chamber is waterproof and corrosion-resistant, and it offers high sensitivity, low power consumption, stable performance, temperature compensation, linear output, resistance to water vapor interference, and no poisoning. Applications include HVAC equipment, air quality monitoring equipment, fresh air systems, air purification equipment, smart home, and schools.
The sensor measures 0-10000 ppm CO2 with selectable ranges and requires 4.5-5.5 V DC, with average current under 20 mA and peak current of 150 mA. It has a 3-minute preheat time, T90 response under 120 seconds, 0-50 C operating temperature, 0-95% RH operating humidity without condensation, and a weight of 5 g. Outputs include UART at 9600 baud with 8 data bits, 1 stop bit, and no parity; PWM with a 1004 ms period; and analog output defaulting to 0.4-2.0 V, configurable from 0-3 V. Zero calibration is at 400 ppm and can be performed manually, by command, or automatically every 26 hours. This manual is written in Chinese.
| Detected gas | Carbon dioxide (CO2) |
|---|---|
| Supply voltage | 4.5-5.5 V DC |
| Average current | Less than 20 mA at 5 V |
| Peak current | 150 mA at 5 V |
| Interface level | 3.3 V, compatible with 5 V |
| Measurement range | 0-10000 ppm, selectable |
| Preheat time | 3 minutes |
| Response time | T90 less than 120 seconds |
| Operating temperature | 0-50 C |
| Operating humidity | 0-95% RH without condensation |
| Accuracy | Plus or minus (50 ppm + 5% of reading) |
| PWM period | 1004 ms plus or minus 5% |
| UART settings | 9600 baud, 8 data bits, 1 stop bit, no parity |
| Analog output | Default 0.4-2.0 V, configurable 0-3 V |
| Zero calibration point | 400 ppm |
| Automatic calibration interval | Every 26 hours |
| Manual zero calibration hold time | Low level for at least 7 seconds |
| Minimum supply current | Not less than 150 mA |
It uses the non-dispersive infrared (NDIR) principle to detect CO2 in air.
It provides UART, analog (DAC), and PWM outputs.
The measurement range is 0-10000 ppm and is selectable.
Set the baud rate to 9600, data bits to 8, stop bits to 1, and parity to none.