Winsen ME4-SO2 SO2 Gas Sensor User Manual

This manual provides usage instructions, technical specifications, and handling precautions for the Winsen ME4-SO2 electrochemical sulfur dioxide gas sensor.


Winsen ME4-SO2 SO2 Gas Sensor User Manual - cover page
Brand
Winsen
Category
Other Electronics
Document type
User Manual
Language
Chinese
Pages
6
File format
PDF
File size
566 KB
Published
17 October, 2026
Updated
17 October, 2026
MD5 checksum
B511FAA2B3FA0E23F417503622C0C35F

About this manual

The document is a user manual for the Winsen ME4-SO2, an electrochemical gas sensor designed to detect sulfur dioxide (SO2) concentrations. It operates on the principle of potentiostatic electrolysis where current generated by oxidation-reduction reactions at the working and counter electrodes corresponds to the SO2 concentration.

The guide details key technical parameters such as a measurement range of 0-20ppm, a response time of less than 30 seconds, and an operating temperature range of -20°C to 50°C. It also includes information on cross-interference from other gases like hydrogen sulfide and carbon monoxide, along with specific installation and maintenance precautions. This manual is written in Chinese.

What's inside

  • Product Description
  • Sensor Characteristics
  • Main Applications
  • Technical Specifications
  • Basic Circuit Diagrams
  • Cross-Interference Characteristics
  • Precautions for Use

Specifications

Detected GasSulfur Dioxide (SO2)
Measurement Range0~20ppm
Maximum Measurement Limit200ppm
Sensitivity(0.8±0.2) µA/ppm
Resolution0.1ppm
Response Time (T90)<=30S
Operating Temperature Range-20℃~50℃
Humidity Range15%~90%RH
Service Life2 years (in air)

Frequently asked questions

What is the recommended load resistance for this sensor?

The manual recommends a load resistance of 10Ω.

How long should the sensor be aged before use?

The aging time must not be less than 48 hours before using the sensor.

Does this sensor respond to hydrogen sulfide gas?

Yes, at a concentration of 50ppm, it shows a response of less than 0.15.

Download PDF (566 KB)

More Winsen Manuals