This manual provides usage instructions, technical specifications, and precautions for the Winsen ME4-O3-E4 electrochemical ozone gas sensor.
The Winsen ME4-O3-E4 is a potentiostatic electrolytic type ozone sensor designed for high precision and sensitivity. It operates by measuring the current generated by oxidation-reduction reactions between ozone and the working and counter electrodes. The sensor features a strong signal level and low zero current, making it suitable for environmental monitoring applications. This manual is written in Chinese.
| Detected Gas | Ozone (O3) |
|---|---|
| Range | 0-20ppm |
| Maximum Measurement Limit | 50ppm |
| Sensitivity | -0.225~-0.525 uA/ppm |
| Resolution | 20ppb |
| Response Time (T90) | ≤60s |
| Recommended Load Resistance | 33-100Ω |
| Operating Temperature Range | -30℃-40℃ |
| Operating Humidity Range | 15%-85% RH |
| Detection Life | 2 years (in air) |
The recommended load resistance is between 33 and 100 ohms.
The sensor must be aged for no less than 48 hours before use.
The maximum measurement limit is 50ppm.