This document is the ME3-O2 electrochemical gas sensor specification of SUNSTAR, which mainly introduces the characteristics of the product, including low power consumption, high accuracy, high sensitivity, wide linear range, strong anti-interference ability, excellent repeatability and stability.
This Chinese product specification describes the SUNSTAR ME3-O2 electrochemical oxygen sensor. The sensor works on the principle of an electrochemical galvanic cell: the gas being measured is reduced at the cathode and oxidized at the anode, producing a current that is proportional to gas concentration and follows Faraday's law, so the concentration can be determined by measuring the current. Listed features are low power consumption, high accuracy, high sensitivity, a wide linear range, strong interference resistance, and good repeatability and stability. Applications given are oxygen detection in industry, underground mines and environmental protection.
The document gives the technical specifications (range, maximum measurement limit, service life, sensitivity, temperature, pressure, response time, humidity, stability, recommended load resistance and repeatability), a basic test circuit diagram, and curves for sensitivity and response/recovery, concentration linearity and temperature characteristics. A precautions section warns against organic solvents, resin encapsulation, immersion in pure oxygen, corrosive gases, blocked or contaminated vent faces, bent or broken pins, and excessive shock, and requires aging of at least 24 hours before use, shorted working and reference electrodes during storage, and no soldering during installation. This manual is written in Chinese.
| Measuring range | 0 to 25 %vol |
|---|---|
| Maximum measurement limit | 30 %vol |
| Service life | 2 years |
| Sensitivity | 0.15 ±0.05 mA (in air) |
| Pressure range | Standard atmosphere ±10% |
| Response time (T90) | ≤ 15 s |
| Humidity range | 0–99 %RH, non-condensing |
| Recommended load resistance | 100 Ω |
| Repeatability | < 2 % of output value |
The measuring range is 0 to 25 %vol, with a maximum measurement limit of 30 %vol.
It must be aged for no less than 24 hours before use, and soldering is not allowed during installation.
A load resistance of 100 Ω is recommended.
The response time (T90) is 15 seconds or less.