Thermo Scientific Small Benchtop Rotors for Thermo Scientific Megafuge 8 / 8R, Sorvall ST 8 / 8R and SL 8 / 8R Centrifuges Instruction Manual
This manual provides instructions for installing, operating, and maintaining small benchtop rotors compatible with specific centrifuges. It includes safety precautions, rotor specifications, loading guidelines, and maintenance procedures. The rotors are designed for laboratory use to separate substances by density and require trained personnel for safe operation.
The document covers rotor installation, balancing, temperature considerations, aerosol-tight applications, cleaning, disinfection, and decontamination. It also includes technical data for various rotor models and service information for Thermo Fisher Scientific.
| Maximum sample density at maximum speed | 1.2 g/cm³ |
|---|
The maximum sample density at maximum speed is 1.2 g/cm³.
Due to air friction, the rotor temperature may rise significantly while spinning. Ventilated units will always lead to temperature increase compared to ambient temperatures. Refrigerated units have limitations in cooling capabilities. Displayed and set temperature can deviate from sample temperature. Sample temperature might exceed critical temperature of your application.