Thermo Fisher Scientific Small Benchtop Rotors for Megafuge 8 / 8R, Sorvall ST 8 / 8R and SL 8 / 8R Centrifuges Instruct
This manual provides instructions for installing, operating, and maintaining small benchtop rotors compatible with specific centrifuges. It includes safety precautions, rotor specifications, loading guidelines, and maintenance procedures. The rotors are designed for laboratory use to separate substances by density and require trained personnel for safe operation.
The document covers rotor installation, balancing, temperature considerations, aerosol-tight applications, cleaning, disinfection, and decontamination. It also includes technical data for various rotor models and service information for Thermo Fisher Scientific.
Thermo Fisher Scientific Sorvall ST 8 specifications
| Max Speed (nmax) | 16000 rpm |
|---|---|
| Max RCF Value at nmax | 24327xg |
| Maximum Kinetic Energy | < 8.12 kJ |
| Noise Level at Maximum Speed | < 58 dB (A) |
| Max Running Time | 99 h 59 min 50 sec |
| Permissible Ambient Temperature during Operation | +2 °C to +35 °C |
Source: Thermo Fisher Scientific Sorvall ST 8 Centrifuge Instruction Manual
All Thermo Fisher Scientific Sorvall ST 8 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is the maximum speed of the Sorvall ST 8 centrifuge?
The maximum speed depends on the rotor, but it can reach up to 16,000 rpm with the TX-150 rotor.
From: Thermo Fisher Scientific Sorvall ST 8 Centrifuge Instruction Manual
What are the permissible ambient temperatures for operation?
The permissible ambient temperature during operation ranges from +2 °C to +35 °C.
From: Thermo Fisher Scientific Sorvall ST 8 Centrifuge Instruction Manual
What is the maximum sample density at maximum speed?
The maximum sample density at maximum speed is 1.2 g/cm³.
What should I do if the rotor temperature increases during operation?
Due to air friction, the rotor temperature may rise significantly while spinning. Ventilated units will always lead to temperature increase compared to ambient temperatures. Refrigerated units have limitations in cooling capabilities. Displayed and set temperature can deviate from sample temperature. Sample temperature might exceed critical temperature of your application.