This document is a release note for Thermo Fisher Scientific Attune™ NxT Software v2.4, which introduces new software features, known software/system issues and fixed software bugs, and provides software installation instructions.
This document provides release notes for Thermo Fisher Scientific Attune™ NxT Software v2.4, detailing new features such as area scaling factors, import/export of workspaces and compensation settings, and support for deep well plates. It also lists known issues with troubleshooting steps and installation instructions. The software is for research use only and not for diagnostic procedures.
The document includes guidance for instrument startup, experiment setup, compensation, and sample acquisition, with specific notes on handling large data files, auto sampler usage, and troubleshooting common errors such as failed baseline tests or software hangs.
Area Scaling Factors (ASF) are now applied from Performance Test (Advanced/System Settings), Import and Export of Workspaces, Import and Export of Compensation Settings, Support for deep well plates up to 2 mLs, User options to set Plot label display name ($PnN vs. $PnS), Implemented load instrument settings and workspace on New Experiment dialog, Support for Blue, Red, Yellow (BRY) instruments, Sample recovery to return unused sample to a tube or preloaded sample in a plate, Drag and Drop of Experiment Explorer Attributes (Instrument Settings, Workspace, Compensation), Filter Configuration for system wide annotation of fluorophores, Implemented context menus on the heat map, Completion of optional FCS file keywords (ie $EXP, $INST, $OP, $CELLS, etc).
Power on the auto sampler before the Attune™ NxT cytometer. If you do not turn the auto sampler on prior to starting the instrument the auto sampler is not recognized.
Remove any filters that are applied to the experiment explorer before creating a new experiment. If the experiment explorer is empty, do not apply a keyword or date filter. If this is done, log out and log in to reset the experiment explorer.
Check that the sample syringe pump (located on the left side of the instrument) is moving. If the pump is not moving, run the unclog function to re-initialize the pump. If the error state continues, power cycle the instrument.