Thermo Fisher Scientific Attune™ NxT Software v2.4 Release Notes

This document is a release note for Thermo Fisher Scientific Attune™ NxT Software v2.4, which introduces new software features, known software/system issues and fixed software bugs, and provides software installation instructions.


Thermo Fisher Scientific Attune™ NxT Software v2.4 Release Notes - cover page
Brand
Thermo Fisher Scientific
Category
Professional medical equipment
Document type
User Manual
Language
English
Pages
13
File format
PDF
File size
526 KB
Published
20 October, 2023
Updated
20 October, 2023
MD5 checksum
9F94B661874C93EA454BBC994A916033

About this manual

This document provides release notes for Thermo Fisher Scientific Attune™ NxT Software v2.4, detailing new features such as area scaling factors, import/export of workspaces and compensation settings, and support for deep well plates. It also lists known issues with troubleshooting steps and installation instructions. The software is for research use only and not for diagnostic procedures.

The document includes guidance for instrument startup, experiment setup, compensation, and sample acquisition, with specific notes on handling large data files, auto sampler usage, and troubleshooting common errors such as failed baseline tests or software hangs.

What's inside

  • New Software Features
  • Known Issues and Workarounds
  • Installation Instructions
  • Instrument Startup Guidance
  • Experiment Setup
  • Compensation Setup
  • Sample Acquisition Guidance

Frequently asked questions

What new features are implemented in Version 2.4?

Area Scaling Factors (ASF) are now applied from Performance Test (Advanced/System Settings), Import and Export of Workspaces, Import and Export of Compensation Settings, Support for deep well plates up to 2 mLs, User options to set Plot label display name ($PnN vs. $PnS), Implemented load instrument settings and workspace on New Experiment dialog, Support for Blue, Red, Yellow (BRY) instruments, Sample recovery to return unused sample to a tube or preloaded sample in a plate, Drag and Drop of Experiment Explorer Attributes (Instrument Settings, Workspace, Compensation), Filter Configuration for system wide annotation of fluorophores, Implemented context menus on the heat map, Completion of optional FCS file keywords (ie $EXP, $INST, $OP, $CELLS, etc).

What should I do if the auto sampler is not recognized during startup?

Power on the auto sampler before the Attune™ NxT cytometer. If you do not turn the auto sampler on prior to starting the instrument the auto sampler is not recognized.

How do I handle a software hang when creating a new experiment?

Remove any filters that are applied to the experiment explorer before creating a new experiment. If the experiment explorer is empty, do not apply a keyword or date filter. If this is done, log out and log in to reset the experiment explorer.

What should I do if the sample syringe pump stalls during a rinse or other fluid function?

Check that the sample syringe pump (located on the left side of the instrument) is moving. If the pump is not moving, run the unclog function to re-initialize the pump. If the error state continues, power cycle the instrument.

Download PDF (526 KB)