This manual introduces the features and operating instructions of REED Instruments R3030/R3100 digital manometer.
This manual provides operating instructions for the R3030 and R3100 digital manometers, including mode options, calibration procedures, and troubleshooting. The R3030 measures 0–30 psi, while the R3100 measures 0–100 psi. Both models feature 11 pressure units, auto power-off, and differential pressure mode. Calibration can be adjusted between 78–82 psi for R3100 and ±24 psi for R3030.
| Range | R3030: 0~30 psi; R3100: 0~100 psi |
|---|---|
| Accuracy | 0.3% of full scale at 25°C |
| Resolution | Model R3030: 0.02 psi; Model R3100: 0.1 psi |
| Response Time | 0.5 seconds |
| Operating Temperature | 0~50°C |
| Power Supply | Single 9V battery |
| Dimensions | 72 x 182 x 30 mm |
| Weight | Approx. 220 g (with battery) |
The display shows 'Err.1' and the meter may be damaged if used beyond its range.
Press and simultaneously before powering on to disable sleep mode.