Key facts
- Brand
- Reed Instruments
- Model
- R3100
- Category
- Other Electronics
- Documents
- 3
- Languages
- EN
- Last added
- 2023-10-19
Start here

REED Instruments R3030/R3100 Digital Manometer Instruction Manual
This manual provides operating instructions for the R3030 and R3100 digital manometers, including mode options, calibration procedures, and troubleshooting. The R3030 measures 0–30 psi, while the R3100 measures 0–100 psi. Both models feature 11 pressure units, auto power-off, and differential pressure mode. Calibration can be adjusted between 78–82 psi for R3100 and ±24 psi for R3030.
Read onlineReed Instruments R3100 specifications
| Range | R3030: 0~30 psi; R3100: 0~100 psi |
|---|---|
| Accuracy | 0.3% of full scale at 25°C |
| Resolution | Model R3030: 0.02 psi; Model R3100: 0.1 psi |
| Response Time | 0.5 seconds |
| Operating Temperature | 0~50°C |
| Power Supply | Single 9V battery |
| Dimensions | 72 x 182 x 30 mm |
| Weight | Approx. 220 g (with battery) |
Source: REED Instruments R3030/R3100 Digital Manometer Instruction Manual
All Reed Instruments R3100 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What happens if the pressure exceeds the range?
The display shows 'Err.1' and the meter may be damaged if used beyond its range.
From: REED Instruments R3030/R3100 Digital Manometer Instruction Manual
How do I disable the auto power-off feature?
Press and simultaneously before powering on to disable sleep mode.
From: REED Instruments R3030/R3100 Digital Manometer Instruction Manual
What happens if the meter is inactive for a while?
The instrument shuts off automatically after approximately 20 minutes of inactivity.
From: Reed Instruments R3030/R3100 Digital Manometer Instruction Manual
How do I disable the auto power-off feature?
Press and hold the power button simultaneously before powering on the device.
From: Reed Instruments R3030/R3100 Digital Manometer Instruction Manual
What happens if the pressure exceeds the range?
The display shows 'Err.1' and the meter may be damaged if used beyond its range.
From: REED Instruments R3030/R3100 Digital Manometer Instruction Manual
How do I disable auto power-off?
Press and simultaneously before powering on to disable sleep mode. An 'n' will appear on screen.
From: REED Instruments R3030/R3100 Digital Manometer Instruction Manual