This is an installation manual for S-2000 StabilizerTM Vibration Isolators and Optical Table
This manual provides step-by-step instructions for installing, leveling, and operating the Newport S-2000 StabilizerTM vibration isolation system. It covers assembly of isolators and table top, air supply requirements, performance adjustments, maintenance, and troubleshooting common issues. The system is designed for vibration-sensitive equipment like interferometers and microscopes.
Safety precautions are emphasized, including grounding the table top and installing earthquake restraints in seismic zones. The manual also includes warranty information and guidance for contacting Newport support for repairs or replacements.
| Maximum Load per Isolator | 2000 pounds |
|---|---|
| Isolator Base Diameter | 10.0 inches |
| Required Air Pressure | Calculated by dividing total load by total isolator diaphragm area (25.9 in² per isolator) plus 5-10 psig |
| Maximum Operating Pressure | 95 psig (6.7 kg/cm2) |
The S-2000 Stabilizer™ isolation system is capable of supporting 2000 pounds per isolator.
The isolator must be mounted so that its axis is not more than 0.5 degrees from vertical.
The required pressure is calculated by dividing the total load by the total isolator diaphragm area (25.9 in² per isolator) plus 5-10 psig.
The maximum operating pressure for the system is 95 psig (6.7 kg/cm2).