Newport S-2000 StabilizerTM Vibration Isolators and Optical Table Installation Manual

This is an installation manual for S-2000 StabilizerTM Vibration Isolators and Optical Table


Newport S-2000 StabilizerTM Vibration Isolators and Optical Table Installation Manual - cover page
Brand
Newport
Document type
User Manual
Language
English
Pages
29
File format
PDF
File size
456 KB
Published
19 October, 2023
Updated
19 October, 2023
MD5 checksum
EEEA02073895EC47FA0B4584256AF546

About this manual

This manual provides step-by-step instructions for installing, leveling, and operating the Newport S-2000 StabilizerTM vibration isolation system. It covers assembly of isolators and table top, air supply requirements, performance adjustments, maintenance, and troubleshooting common issues. The system is designed for vibration-sensitive equipment like interferometers and microscopes.

Safety precautions are emphasized, including grounding the table top and installing earthquake restraints in seismic zones. The manual also includes warranty information and guidance for contacting Newport support for repairs or replacements.

What's inside

  • General Information
  • Assembly
  • Operation
  • Troubleshooting
  • Factory Service

Specifications

Maximum Load per Isolator2000 pounds
Isolator Base Diameter10.0 inches
Required Air PressureCalculated by dividing total load by total isolator diaphragm area (25.9 in² per isolator) plus 5-10 psig
Maximum Operating Pressure95 psig (6.7 kg/cm2)

Frequently asked questions

What is the maximum load capacity per isolator?

The S-2000 Stabilizer™ isolation system is capable of supporting 2000 pounds per isolator.

How should the isolators be mounted for proper function?

The isolator must be mounted so that its axis is not more than 0.5 degrees from vertical.

What is the required air pressure for operation?

The required pressure is calculated by dividing the total load by the total isolator diaphragm area (25.9 in² per isolator) plus 5-10 psig.

What is the maximum operating pressure for the system?

The maximum operating pressure for the system is 95 psig (6.7 kg/cm2).