FAFNIR LS 300 and LS 500 H SIL Safety Manual

This document introduces the safety manual for FAFNIR's LS 300 and LS 500 H SIL sensors, which are suitable for areas that require safety assessment according to EN 61508


FAFNIR LS 300 and LS 500 H SIL Safety Manual - cover page
Brand
FAFNIR
Category
Industrial Supply
Document type
Service Manual
Language
English
Pages
3
File format
PDF
File size
121 KB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
69BBECC755A26ED308033711C26735FC

About this manual

This document serves as a safety manual for the FAFNIR LS 300 level detector and LS 500 H SIL measuring transducer series. It details the application of these devices in safety-related subsystems requiring SIL 2 compliance according to EN 61508, specifically for overfill and dry-running protection.

What's inside

  • Range of application
  • Standards
  • Instructions for safe use
  • Assembling and dismantling
  • Installation
  • Adjustment
  • Putting into service
  • Maintenance
  • Safety Ratings
  • Additional requirements for software elements

Specifications

Response time (dry-run protection)approximately 20 s
Vibration test range10 Hz to 55 Hz and 2 g
Mean time to restoration8 h
Interval of the retest1 year

Frequently asked questions

What is the minimum resistance value required for the level sensor to detect short circuits?

The resistance value must be less than 30 Ω (or < 2.5 Ω with firmware version 1.0.1.255).

How often should the functionality of the system be checked?

The functionality should be checked at appropriate intervals, but at least once a year.

What happens if the test button of a surfaced level detector is pressed?

It switches to the "immersed" state after less than a second and the sensor relay is deactivated.