The safety manual introduces the product's technical characteristics, including the product's scope of application, standards, safety instructions, etc.
This document is a safety manual for the FAFNIR LS 300 level detector and LS 500 H SIL measurement transducer. It is designed for use in safety subsystems compliant with the EN 61508 SIL 2 standard, specifically for anti-overflow and dry-run protection applications. This manual is written in French.
| Safety Integrity Level | SIL 2 |
|---|---|
| Mean Time To Repair | 8 h |
| Safety Failure Rate (λSD) | 1,92E-06 |
| Safety Failure Rate (λSU) | 4,44E-08 |
| Safety Failure Rate (λDD) | 2,12E-06 |
| Safety Failure Rate (λDU) | 6,83E-07 |
| Probability of Dangerous Failure per Hour (PFH) | 6,83E-07 |
| Safety Test Interval | 1 an |
| Architecture | 1001 |
The maximum reaction time is 2 seconds.
The typical reaction time is approximately 20 seconds at room temperature.
The system should be checked at appropriate intervals, and at least once a year.