About this manual

This document is a safety manual for the FAFNIR LS 300 level detector and LS 500 H SIL measurement transducer. It is designed for use in safety subsystems compliant with the EN 61508 SIL 2 standard, specifically for anti-overflow and dry-run protection applications. This manual is written in French.

What's inside

  • Domaine d'application
  • Normes
  • Instructions pour effectuer sans risques
  • Le montage et le démontage
  • La mise en service
  • La maintenance
  • Caractéristiques techniques relatives à la sécurité

Specifications

Safety Integrity LevelSIL 2
Mean Time To Repair8 h
Safety Failure Rate (λSD)1,92E-06
Safety Failure Rate (λSU)4,44E-08
Safety Failure Rate (λDD)2,12E-06
Safety Failure Rate (λDU)6,83E-07
Probability of Dangerous Failure per Hour (PFH)6,83E-07
Safety Test Interval1 an
Architecture1001

Frequently asked questions

What is the maximum reaction time for the LS 300 as an anti-overflow safety?

The maximum reaction time is 2 seconds.

What is the typical reaction time for the LS 300 as a dry-run protection?

The typical reaction time is approximately 20 seconds at room temperature.

How often should the system be checked for proper operation?

The system should be checked at appropriate intervals, and at least once a year.