FAFNIR LS 300 Level Detector Safety Manual
This safety manual covers the proper use, installation, and maintenance of FAFNIR LS 300 level detectors and LS 500 H SIL measurement converters for SIL 2 safety applications. It includes safety standards, wiring diagrams, diagnostic procedures, and error codes for troubleshooting. The device is designed to prevent overfilling or dry-running in industrial environments.
The manual specifies safe operating conditions, including vibration resistance up to 2g and 55Hz, and provides instructions for firmware version reading via MODBUS-RTU protocol. It also details error handling, including SIL error codes and reset procedures. This manual is written in Spanish.
FAFNIR LS 300 specifications
| Safety Integrity Level | SIL 2 |
|---|---|
| Hardware Fault Tolerance | HFT = 0 |
| Mean Time to Recovery | 8 h |
| Repetition Test Interval | 1 año |
| Response Time (Overfill Protection) | 2 segundos |
| Response Time (Dry-Run Protection) | 20 segundos |
All FAFNIR LS 300 documents by type
Disassembly, diagnostics and part numbers, for repair.
Questions answered in these manuals
How do I read the firmware version via MODBUS-RTU?
Use function code 03 (Read Holding Registers) with addresses 0x0040 and 0x0041 to read the version number.
What is the SIL error code for memory failure?
Memory failure is indicated by intermittent code 1.
What safety standard does this device comply with?
The device complies with EN 61508:2010 for functional safety of electrical/electronic/electronic programmable systems, with SIL 2 rating.
From: EN 61508 Safety Manual
How is the device diagnosed for SIL errors?
SIL errors are indicated by a blinking LED "Error" (from firmware 1.2.0.255 and hardware 1.1.0.255), with specific codes for memory, checksum, relay, resistance scanner, transistor scanner, and isolation faults.
From: EN 61508 Safety Manual
What is the minimum resistance value required for the level sensor to detect short circuits?
The resistance value must be less than 30 Ω (or < 2.5 Ω with firmware version 1.0.1.255).
How often should the functionality of the system be checked?
The functionality should be checked at appropriate intervals, but at least once a year.