This file is the user manual for Thermo Scientific Orion Fluoride Ion Selective Electrode
This manual provides instructions for preparing, operating, and maintaining the fluoride ion selective electrode (ISE). It includes details on required equipment, electrode setup, calibration methods, and troubleshooting. The electrode is designed for accurate fluoride measurement in aqueous solutions, with EPA-approved methods for drinking water and wastewater. It also covers electrode storage, analytical techniques, and technical support resources.
The filling solution level must be at least one inch above the level of the sample in the beaker.
There should be a 54 to 60 mV difference between the two millivolt readings.
Low-Level TISAB is for use when measuring samples containing less than 0.4 ppm fluoride.