Thermo Fisher Scientific Orion Fluoride Ion Selective Electrode 254792-001 User Manual

This manual describes the use of the Orion Fluoride Ion Selective Electrode 254792-001. The electrode is used for the determination of fluoride ion in water, wastewater, and food.


Thermo Fisher Scientific Orion Fluoride Ion Selective Electrode 254792-001 User Manual - cover page
Brand
Thermo Fisher Scientific
Category
Health Care
Document type
User Manual
Language
English
Pages
41
File format
PDF
File size
351 KB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
BC54C95B2AC10B7D069111B3F6DA925C

About this manual

This manual provides instructions for preparing, operating, and maintaining the Thermo Fisher Scientific Orion Fluoride Ion Selective Electrode 254792-001. It includes details on required equipment, electrode setup, sample preparation, calibration methods, and troubleshooting. The electrode is designed for accurate fluoride ion measurement in aqueous solutions and complies with EPA-approved test procedures.

The manual also covers electrode storage, analytical techniques including direct calibration and titrations, and special buffers like TISAB II and TISAB IV for different sample conditions. It specifies compatible meters and reference electrodes, and includes electrode filling instructions and slope verification procedures.

What's inside

  • Introduction
  • Required Equipment
  • Electrode Setup
  • Electrode Preparation
  • Checking Electrode Operation (Slope)
  • Sample Requirements
  • Electrode Storage
  • Analytical Techniques
  • Electrode Characteristics
  • Troubleshooting

Specifications

Electrode TypeFluoride Ionplus® Sure-Flow® Combination ISE
Filling SolutionOptimum Results A filling solution, Cat. No. 900061
Electrode Slope54 to 60 mV

Frequently asked questions

What is the electrode slope and how is it checked?

The electrode slope is defined as the change in millivolts observed with every tenfold change in concentration. To check it, prepare the electrode, connect it to a meter, and measure the potential difference between 0.1 M NaF and 100 ppm fluoride standards. The difference should be 54 to 60 mV at 20–25°C.

What are the compatible meters for this electrode?

Compatible meters include the Orion Star A214 pH/ISE meter, Orion Dual Star two channel pH/ISE meter, and Orion Versa Star Pro pH/ISE meter. The electrode can also be used on any ISE meter with a BNC connection.