This manual describes the use of the Orion Fluoride Ion Selective Electrode 254792-001. The electrode is used for the determination of fluoride ion in water, wastewater, and food.
This manual provides instructions for preparing, operating, and maintaining the Thermo Fisher Scientific Orion Fluoride Ion Selective Electrode 254792-001. It includes details on required equipment, electrode setup, sample preparation, calibration methods, and troubleshooting. The electrode is designed for accurate fluoride ion measurement in aqueous solutions and complies with EPA-approved test procedures.
The manual also covers electrode storage, analytical techniques including direct calibration and titrations, and special buffers like TISAB II and TISAB IV for different sample conditions. It specifies compatible meters and reference electrodes, and includes electrode filling instructions and slope verification procedures.
| Electrode Type | Fluoride Ionplus® Sure-Flow® Combination ISE |
|---|---|
| Filling Solution | Optimum Results A filling solution, Cat. No. 900061 |
| Electrode Slope | 54 to 60 mV |
The electrode slope is defined as the change in millivolts observed with every tenfold change in concentration. To check it, prepare the electrode, connect it to a meter, and measure the potential difference between 0.1 M NaF and 100 ppm fluoride standards. The difference should be 54 to 60 mV at 20–25°C.
Compatible meters include the Orion Star A214 pH/ISE meter, Orion Dual Star two channel pH/ISE meter, and Orion Versa Star Pro pH/ISE meter. The electrode can also be used on any ISE meter with a BNC connection.