Digi-Sense TC9100 and TC9600 Advanced Temperature Controllers Calibration Manual

This document is a calibration manual for Digi-Sense® TC9100 and TC9600 Advanced Temperature Controllers. It explains how to calibrate the temperature controller and provides the range of calibration temperature points.


Digi-Sense TC9100 and TC9600 Advanced Temperature Controllers Calibration Manual - cover page
Brand
Digi-Sense
Document type
Service Manual
Language
English
Pages
5
File format
PDF
File size
719 KB
Published
18 October, 2023
Updated
18 October, 2023
Models
TC9100, TC9600
MD5 checksum
64660AF1ADE4D9361A79492206C791A8

About this manual

This manual details how to calibrate Digi-Sense TC9100 and TC9600 temperature controllers using built-in calibration gateways. It covers full-range and segmented calibration methods for various sensor types including thermocouples, platinum RTDs, and thermistors. Each unit comes with preloaded default calibration points for accurate readings across the full temperature range.

Calibration requires specific equipment such as a Fluke 753 or equivalent for thermocouples and RTDs, and an IET decade box for thermistors. The process involves setting six temperature points, confirming calibration, and saving or discarding changes.

What's inside

  • Introduction
  • Full-Range Calibration
  • Segmented Calibration
  • Default Calibration Points
  • Calibration Mode Setup
  • Equipment Requirements

Specifications

Calibration PointsSix temperature points
Thermocouple Accuracy+/-0.1% of span or +/-oC for types J, K, T, E, N; +/- 0.2% of span or +/-2oC for types B,R,S
Platinum RTD Accuracy+/-0.1% of span
Thermistor Accuracy+/-0.4% of span or +/-0.5oC

Frequently asked questions

What equipment is needed to calibrate the controller?

Fluke 753 or equivalent for thermocouples and RTDs, and an IET decade box for thermistors.

How many calibration points can be set?

Six temperature points can be set for calibration.

What are the default calibration points for thermocouples?

Default points vary by type: e.g., Type J: -185, -85, 0, 230, 660, 995 °C.

Can I change the default calibration points?

Yes, but it is not recommended for full-range calibration due to nonlinearity of temperature probes.