This manual describes the measurement requirements, usage and maintenance methods of SEFRAM 9835 digital sound level meter.
The SEFRAM 9835 is a Class 2 digital sound level meter designed to meet IEC61672-1 standards. It features both AC and DC signal outputs suitable for frequency analyzers, recorders, and FFT analyzers. The device includes a ½ inch electret condenser microphone and offers A and C frequency weighting options.
| Standard Applied | IEC61672 -1 Class2 |
|---|---|
| Frequency Range | 20Hz ~ 8KHz |
| Measuring level Range | 32 ~ 130dB |
| Frequency Weighting | A / C |
| Microphone Type | ½ inch electret condenser microphone |
| Digital Display Resolution | 0.1dB |
| Accuracy | ±1.4dB (under reference conditions) |
| Dynamic Range | 60 dB |
The recommended recalibration cycle for the meter is 1 year.
Approximately 60 hours of operation time can be expected with an alkaline battery.
The display will show 'OVER' when the input is more than the upper limit of the range.