This document is the user manual of POINT FOUR™ TRACKER Portable TGP Meter, which introduces the composition, key functions, usage methods, maintenance and maintenance of the meter.
The Pentair Point Four Tracker is a portable Total Gas Pressure (TGP) meter designed to measure total dissolved gas pressure (TDGP) and temperature. The device consists of a handheld meter and a TGP probe, featuring a display for primary and secondary parameters such as Delta P, % saturation, barometric pressure, and temperature. It includes components like a charger cable, data transfer cable, and a TGP pressure test syringe kit.
| Measured Parameters | Total Dissolved Gas Pressure (TGP), Temperature, Barometric Pressure (BP) |
|---|---|
| Probe Length | 6 m / 16.4 ft (optional up to 300 m/ 984 ft) |
| Calibration Frequency | Once every 3 weeks |
Calibration should be undertaken if sensor readings appear incorrect, and the probe must be removed from in-situ measurement and dried completely prior to calibrating.
The probe should be dried completely and calibrated once every 3 weeks, or more frequently if used in water above 20 degrees C.
The probe must be left in air for a minimum of 10 minutes to allow the pressure to equilibrate with atmospheric pressure.