Tektronix MSO3000/DPO3000 Series Mixed Signal Oscilloscopes Data Sheet

This document describes the features and performance specifications of the Mixed Signal Oscilloscopes MSO3000 Series and DPO3000 Series. The series includes models with various bandwidths and channel configurations, high sample rates, and large record lengths. The products also have user-friendly features and interfaces, such as Wave Inspector® Controls, automated measurements, USB and Ethernet ports. Additionally, there are optional serial triggering and analysis options, as well as mixed signal design and analysis capabilities.


Tektronix MSO3000/DPO3000 Series Mixed Signal Oscilloscopes Data Sheet - cover page
Brand
Tektronix
Category
Oscillator
Document type
Datasheet
Language
English
Pages
20
File format
PDF
File size
2.1 MB
Published
02 February, 2023
Updated
28 September, 2023
MD5 checksum
7CA9F423BBF857AF8A8D31B5AB00419F

About this manual

The Tektronix MSO3000 Series Mixed Signal Oscilloscope (MSO) and Digital Phosphor Oscilloscope (DPO) series are designed to analyze up to 20 analog and digital signals with a single instrument. These oscilloscopes feature bandwidths up to 500 MHz, sample rates of 2.5 GS/s on all channels, and a standard record length of 5 Megapoints.

What's inside

  • Features & Benefits
  • Key Performance Specifications
  • Wave Inspector Controls
  • Automated Measurements
  • TekVP I Probe Interface
  • Connectivity
  • Mixed Signal Design and Analysis
  • Power Analysis
  • HDTV and Custom Video Analysis

Specifications

Bandwidth Models500, 300, 100 MHz
Analog Channels2 and 4 Analog Channel Models
Digital Channels (MSO Series)16 Digital Channels
Sample Rate2.5 GS/s on All Channels
Record Length5 Megapoint Record Length on All Channels
Maximum Waveform Capture Rate>50,000 wfm/s
Automated Measurements29 Automated Measurements
Display Size9 in. (229 mm)
Depth5.8 in. (147 mm)

Frequently asked questions

What is the maximum waveform capture rate?

The oscilloscope offers a maximum waveform capture rate of over 50,000 waveforms per second.

Which serial buses are supported for automated triggering and decode?

The MSO/DPO3000 series supports I2C, SPI, CAN, LIN, RS-232/422/485/UART, and I2S/LJ/RJ/TDM.

What is the fine timing resolution provided by MagniVu?

MagniVu high-speed acquisition provides 121.2 ps fine timing resolution on digital channels.

Download PDF (2.1 MB)