Hanna Instruments HI-96742 Ion Selective Meter Instruction Manual

This document is the instruction manual for the HI 96742 Ion Selective Meter, describing the features and characteristics of the product.


Hanna Instruments HI-96742 Ion Selective Meter Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
747 KB
Published
10 July, 2012
Updated
28 September, 2023
MD5 checksum
00FB588B5B1E16D0181D171787F18317

About this manual

This document is an instruction manual for the Hanna Instruments HI-96742 Ion Selective Meter. The device measures Iron Low Range (LR) and Manganese LR concentrations using specific reagents and methods (TPTZ and PAN). It features a liquid crystal display, automatic shut-off, and includes instructions for zeroing, reading samples, and performing validation or calibration procedures using Hanna CAL CHECK standards.

The manual details the functional description of the keys, including RANGE/GLP/A, CAL CHECK, ZERO/CFM, READ/M>/TIMER, and ON/OFF. It provides step-by-step guides for measurement procedures, error handling (such as Light High/Low or Battery low), and troubleshooting common issues like inverted cuvettes or over-range conditions.

What's inside

  • Preliminary examination
  • Technical specifications
  • Functional description
  • Errors and warnings
  • Measurement procedure
  • Validation procedures
  • Calibration procedures

Specifications

Range (Iron LR)0.00 to 1.60 mg/l
Range (Manganese LR)0 to 300 µg/t
Accuracy (Iron LR)+/-0.01 mg/l +5% of reading @ 25°C
Accuracy (Manganese LR)+/-2 µg/L +/-3% of reading @ 25°C
Resolution (Iron LR)0.01 mg/l
Resolution (Manganese LR)1 µg/t
Light SourceSilicon Photocell with narrow band interference filter @ 650 nm
Operating Temperature0 to 50°C (82 to 122°F)
Dimensions192 x 104x 69 mm

Frequently asked questions

What is the method used for Iron LR measurement?

The instrument uses the TPTZ method, where the reaction between iron and the reagent causes a violet tint in the sample.

What is the method used for Manganese LR measurement?

The instrument uses the PAN Method, where the reaction between manganese and the reagents causes an orange tint in the sample.

What should be done if a 'Light High' error appears?

There is too much light to perform a measurement; you should check the preparation of the zero cuvette.

What does an 'Over Range' error indicate?

The concentration of the sample is beyond the programmed range, and you should dilute the sample and re-run the test.

Download PDF (747 KB)