Hanna Instruments HI 95707C Nitrite Low Range ISM Instruction Manual

This manual provides you with the necessary information for the correct use of the instrument. Please read it carefully before using the meter.


Hanna Instruments HI 95707C Nitrite Low Range ISM Instruction Manual - cover page
Brand
Hanna Instruments
Category
Analyzers
Document type
User Manual
Language
English
Pages
10
File format
PDF
File size
870 KB
Published
10 July, 2012
Updated
30 September, 2023
MD5 checksum
C2C7641A6F1A8E3E865F64B8EEA98F03

About this manual

The HI 95707 is an auto-diagnostic portable microprocessor meter designed to measure Nitrogen-Nitrite (NO2--N) content in drinking, surface, saline water, and domestic or industrial wastes. The instrument features an advanced optical system with a tungsten lamp and narrow band interference filter for accurate readings. It includes a CAL CHECK validation function using NIST traceable standards to verify performance at any time.

The meter operates by measuring the pink tint caused by the reaction between nitrite and powder reagents supplied in packets. Users can perform preliminary examinations, calibration procedures, and battery replacements as detailed in this guide. The device is splash-proof and includes an auto-shut-off feature.

What's inside

  • Preliminary Examination
  • General Description
  • Abbreviations
  • Specifications
  • Precision and Accuracy
  • Principle of Operation
  • Functional Description
  • Guide to Display Codes
  • Measurement Procedure
  • Calibration Procedure

Specifications

Range0.000 to 0.600 mg/L (as NO2--N)
Resolution0.001 mg/L
Precision+ 0.001 mg/L @ 0.100 mg/L
Light SourceTungsten lamp with narrow band interference filter @ 525 nm
Light DetectorSilicon Photocell
MethodAdaptation of an EPA approved method
Environment Temperature0 to 50°C (32 to 122°F)
Battery Type1x9 volt

Frequently asked questions

What is the measurement range of the HI 95707C?

The instrument measures Nitrogen-Nitrite content in the range of 0.000 to 0.600 mg/L.

How do I restore the factory calibration?

Enter calibration mode by holding CAL CHECK for three seconds, then hold ZERO for 10 seconds until 'F.CAL' and '707' appear.

What type of battery does the meter use?

The instrument uses a single 9V battery.

Download PDF (870 KB)