Hanna Instruments HI 95711C Free & Total Chlorine ISM Instruction Manual

This is an instruction manual that describes the use of HI 95711C Free & Total Chlorine ISM 24 produced by Hanna Instruments Inc.


Hanna Instruments HI 95711C Free & Total Chlorine ISM Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
12
File format
PDF
File size
911 KB
Published
10 July, 2012
Updated
30 September, 2023
MD5 checksum
F0CFF930A52D4B41E47BAD80A270845C

About this manual

The HI 95711C is an auto-diagnostic portable microprocessor meter designed to measure free and total chlorine content in water samples within a range of 0.00 to 5.00 mg/L (ppm). The instrument features an advanced optical system based on a special tungsten lamp and a narrow band interference filter, ensuring accurate and repeatable readings. It includes the CAL CHECK validation function using NIST traceable standards to verify performance at any time.

The manual provides instructions for preliminary examination, operational procedures, measurement and validation steps, calibration, battery replacement, and maintenance. It also lists required reagents and accessories, such as specific cuvet sets and cleaning solutions compatible with the meter.

What's inside

  • Preliminary Examination
  • General Description
  • Specifications
  • Principle of Operation
  • Functional Description
  • Guide to Display Codes
  • Operational Guide
  • Measurement Procedure
  • Calibration Procedure
  • Battery Replacement

Specifications

Range Free Cl20.00 to 5.00 mg/L
Range Total Cl20.00 to 5.00 mg/L
Resolution0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L
Precision+0.02 mg/L @ 1.00 mg/L
Light SourceTungsten lamp with narrow band interference filter @ 525 nm
Battery Type1x9 volt
Auto-Shut off (measurement mode)After 10 minutes of non-use
Dimensions180 x 83 x 46 mm (7.1 x 3.3 x 1.8")

Frequently asked questions

What is the measurement range for free and total chlorine?

The meter measures free and total chlorine in water samples in the 0.00 to 5.00 mg/L (ppm) range.

How do I restore factory calibration?

Enter calibration mode by holding CAL CHECK for three seconds, then hold ZERO/SET for 10 seconds until the display shows 'F.CAL' and the parameter code.

What reagents are required for testing?

HI 93701-0 is required for Free Chlorine tests, and HI 93711-0 is required for Total Chlorine tests.