About this manual

The HI 95729 is an auto-diagnostic portable microprocessor meter designed to measure fluoride levels in drinking, surface, and waste waters within a range of 0.00 to 2.00 mg/L (ppm). The instrument utilizes an advanced optical system based on a tungsten lamp and a narrow band interference filter at 575 nm, employing the SPADNS method adapted from EPA Method 340.1.

The device features a positive-locking cuvet system to ensure consistent positioning for accurate readings. It includes factory calibration with an option to restore it via the CAL CHECK function and utilizes a 9V battery. The meter is splash-proof and equipped with display codes to aid routine operation, including auto-shutoff after periods of non-use.

What's inside

  • Preliminary Examination
  • General Description
  • Abbreviations
  • Specifications
  • Precision and Accuracy
  • Principle of Operation
  • Functional Description
  • Guide to Display Codes
  • Measurement Procedure
  • Calibration Procedure

Specifications

Range0.00 to 2.00 mg/L
Resolution0.01 mg/L
Precision+/-0.03 mg/L @ 1.00 mg/L
Light SourceTungsten lamp with narrow band interference filter @ 575 nm
Light DetectorSilicon Photocell
Environment0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery Type1x9 volt
Dimensions180 x 83 x 46 mm (7.1 x 3.3 x 1.8")

Frequently asked questions

What is the measurement range of the HI 95729C?

The instrument measures fluoride content in the 0.00 to 2.00 mg/L (ppm) range.

What is the measurement range for this meter?

The HI 95729 measures fluoride content in the 0.00 to 2.00 mg/L (ppm) range.

What water types can this instrument measure?

The meter is designed for measuring fluoride in drinking, surface and waste waters.

How long does the auto-shut off feature last?

The instrument turns off after 10 minutes of non use or after 1 hour if left in calibration mode.