Hanna Instruments HI 95761C Total Chlorine Meter Instruction Manual

This is a total chlorine meter for the analysis of trace Total Chlorine concentrations.


Hanna Instruments HI 95761C Total Chlorine Meter Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
10
File format
PDF
File size
879 KB
Published
10 July, 2012
Updated
30 September, 2023
MD5 checksum
87F4D9972BFBBAA55003B872BFE668B9

About this manual

The HI 95761C is an auto-diagnostic portable microprocessor meter designed for analyzing trace total chlorine concentrations in water samples. It features an advanced optical system with a tungsten lamp and narrow band interference filter, ensuring accurate and repeatable readings. The device includes a CAL CHECK validation function and is splash waterproof for field applications.

What's inside

  • Preliminary Examination
  • General Description
  • Specifications
  • Precision and Accuracy
  • Principle of Operation
  • Functional Description
  • Guide to Display Codes
  • Measurement Procedure
  • Calibration Procedure
  • Battery Replacement

Specifications

Range0.000 to 0.500 mg/L
Resolution0.001 mg/L
Precision+0.004 mg/L @ 0.200 mg/L
Light SourceTungsten lamp with narrow band interference filter @ 525 nm
Light DetectorSilicon Photocell
MethodAdaptation of the USEPA METHOD 330.5
Environment0 to 50°C (32 to 122°F); max 95% RH non-condensing
Auto-Shut offAfter 10' of non-use in measurement mode; after 1 hour of non-use in calibration mode

Frequently asked questions

How do I restore factory calibration?

Hold CAL CHECK for three seconds, then hold ZERO for 10 seconds until 'F.CAL' and '761' appear.

What is the warranty period?

The instrument is warranted for two years against defects in workmanship and materials.

What reagents are required for testing?

The required reagent is Total Chlorine LR Reagent, supplied as HI 95761-0.

Download PDF (879 KB)