Hanna Instruments HI 504920 Remote User Interface pH Meter Instruction Manual

HI 504920 Remote User Interface pH Meter is a portable pH/ORP/temperature meter for use primarily with Hanna HI 504910 Digital Transmitter for two different applications: it can be a reference pH meter for calibrating the HI 504910 Digital Transmitter, or it can be set as user interface for the HI 504910 as well. All pH measurements are automatically compensated for temperature (ATC). The instrument housing is made of rugged, lightweight material, making it truly portable. Five memorized buffers (4.01, 6.86, 7.01, 9.18 and 10.01 pH) and wrong buffer recognition technology make calibration simple and error free. One or two-point calibration can be performed. An user friendly menu navigation makes operation easy. The large backlit LCD provides clear, easy-to-read readings.


Hanna Instruments HI 504920 Remote User Interface pH Meter Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
15
File format
PDF
File size
738 KB
Published
10 July, 2012
Updated
30 September, 2023
Model
HI 504
MD5 checksum
97DB56C95243A8074F99DE0BC41128DA

About this manual

The Hanna Instruments HI 504920 is a complete portable pH/ORP/temperature meter designed to be used with the HI 504910 Digital Transmitter. It features an automatic temperature compensation (ATC) system, five memorized buffers for calibration, and a rugged IP67-rated housing suitable for harsh industrial conditions.

The device allows users to perform one or two-point calibrations using pH buffers such as 4.01, 6.86, 7.01, 9.18, and 10.07. It includes a Battery Error Preventing System (BEPS) that warns of low battery levels while continuing to measure correctly until the batteries are too weak. The meter can be powered by four AA batteries or an external 12 VDC adapter.

What's inside

  • Preliminary Examination
  • General Description
  • Functional Description
  • Specifications
  • Initial Preparation
  • Calibration
  • Good Laboratory Practice (GLP)
  • Battery Replacement

Specifications

Range pH-4.00 to 19.99 pH
Resolution pH0.07 pH
Accuracy pH+0.01 pH @20°C/68°F
Temperature Range-20 to 120 °C
IP RatingIP 67
Power Supply (Battery)4x1.5V AA batteries

Frequently asked questions

Which pH buffers are memorized for calibration?

The meter has five memorized buffers: 4.01, 6.86, 7.01, 9.18 and 10.07 pH.

Can the meter be used with external power?

Yes, for long term field and lab applications, this meter can be connected to a 12 VDC adapter.

What does the "WRONG" message indicate during calibration?

It alerts the user if a wrong solution or electrode has been used or if the buffer is polluted.