Hanna Instruments HI 706 Phosphorus High Range Meter Instruction Manual

Phosphorus High Range Analyzer


Hanna Instruments HI 706 Phosphorus High Range Meter Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
579 KB
Published
10 July, 2012
Updated
30 September, 2023
MD5 checksum
FD0A6BC74AFD015150C3D2FB7909B98A

About this manual

The document is an instruction manual for the Hanna Instruments HI 706, a portable photometric analyzer designed to measure reactive phosphorus in water and wastewater. The instrument utilizes the Heteropoly-molybdenum Blue method, where the reaction between reactive phosphorus and reagent causes a blue tint in the sample.

What's inside

  • Preliminary examination
  • Functional description
  • Errors and warnings
  • Technical specifications
  • Measurement procedure
  • Accessories
  • Interference
  • Battery management

Specifications

Range0.0 to 15.0 ppm
Resolution0.1 ppm
Accuracy0.3 ppm +5 % of reading @ 25 °C / 77
Light SourceLight Emitting Diode @ 525 nm
Light DetectorSilicon Photocell
Environment Temperature0 to 50 °C (32 to 122 °F)
Relative Humiditymax 95% RH non-condensing
Battery Type1 x 1.5 V AAA Battery
Dimensions81.5 x 61 x 37.5 mm (3.2 x 2.4x1.5")
Weight64g (2.25 oz)

Frequently asked questions

What method does the HI 706 use for measurement?

It uses the Heteropoly-molybdenum Blue method from Standard Methods for the Examination of Water and Wastewater, 18th edition.

What happens if there is too much light?

The instrument displays a warning indicating that there is too much light to perform the measurement.

How long does it take for the battery to shut down automatically?

The instrument shuts off after 10 minutes of non-use.