Hanna Instruments HI 7101/HI 7102 RH Calibration Kits Instruction Manual

This is a user manual of Hanna HI 7101 and HI 7102 RH calibration kits. These kits include two thermally isolated chambers, each one equipped with a cap and three bottles containing the appropriate precalibrated saturated salts to produce a known RH value: two 15-gram bottles of Lithium Chloride (LiCl) and one 33-gram bottle of Sodium Chloride (NaCl).


Hanna Instruments HI 7101/HI 7102 RH Calibration Kits Instruction Manual - cover page
Brand
Hanna Instruments
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
140 KB
Published
10 July, 2012
Updated
30 September, 2023
MD5 checksum
21950FAC43AA31BE105F38A67A1C0029

About this manual

The document provides instructions for using the HI 7101 and HI 7102 portable humidity calibration chambers. These kits include thermally isolated chambers equipped with precalibrated saturated salts (Lithium Chloride and Sodium Chloride) to produce known Relative Humidity (RH) values of approximately 11.1% and 75.4%. The manual specifies compatible Hanna Instruments hygrometer models for each kit.

What's inside

  • Preliminary Examination
  • General Description
  • Preparing the Calibration Solutions
  • Calibration Procedure
  • Warranty

Frequently asked questions

What are the dimensions of the HI 7101 and HI 7102 calibration kits?

The portable chambers measure only 190 x 115 x 110 mm.

How long does the calibration kit need to stabilize before use?

The calibration kit needs 4 hours for stabilization.

What is the acceptable reading range when adjusting the low humidity trimmer to zero?

A value of 0.0% is required, though a reading between 0.0% and 1.0% RH is also acceptable.