Agilent 6890 Gas Chromatograph OQ/PV Procedure Manual

Agilent 6890 Gas Chromatography Headspace Static Sampling OQ/PV Procedures


Agilent 6890 Gas Chromatograph OQ/PV Procedure Manual - cover page
Brand
Agilent
Document type
Service Manual
Language
French
Pages
2
File format
PDF
File size
39 KB
Published
06 July, 2012
Updated
30 September, 2023
MD5 checksum
F4D05D9D31FE8D6D2DEBC609C293C985

About this manual

The document outlines a testing procedure to prove that an Agilent 6890 Gas Chromatograph coupled with a static headspace sampler functions according to specifications. The process involves internal tests and analytical methods using certified samples to evaluate system performance, typically completed within one day.

Prerequisites for the test include recommended injector maintenance, installation of ChemStation version A.05.XX or later (certified), and specific hardware configurations involving a G1580A GC with FID detector and either a G1289A or G1290A headspace sampler. This manual is written in French.

What's inside

  • Conditions préalables
  • Configuration matérielle supportée
  • Mesure des débits
  • Précision de la température
  • Intégrité de la pression d'injecteur
  • Bruit et instabilité du détecteur
  • Test chimique: surface absolue
  • Test chimique: effet mémoire

Specifications

Software Version RequirementA.05.XX or later
Supported GC ModelG1580A
Supported Headspace Sampler ModelsG1289A or G1290A

Frequently asked questions

How long does the OQ/PV procedure typically take?

The procedure can generally be completed in one day.

What is the purpose of the injector pressure integrity test?

This test verifies that the injector maintains pressure according to Agilent 6890 specifications.

Which standard is used for the detector noise and instability test?

The test is performed according to ASTM E 685-93 guidelines.