Agilent 6890 Gas Chromatography Headspace Static Sampling OQ/PV Procedures
The document outlines a testing procedure to prove that an Agilent 6890 Gas Chromatograph coupled with a static headspace sampler functions according to specifications. The process involves internal tests and analytical methods using certified samples to evaluate system performance, typically completed within one day.
Prerequisites for the test include recommended injector maintenance, installation of ChemStation version A.05.XX or later (certified), and specific hardware configurations involving a G1580A GC with FID detector and either a G1289A or G1290A headspace sampler. This manual is written in French.
| Software Version Requirement | A.05.XX or later |
|---|---|
| Supported GC Model | G1580A |
| Supported Headspace Sampler Models | G1289A or G1290A |
The procedure can generally be completed in one day.
This test verifies that the injector maintains pressure according to Agilent 6890 specifications.
The test is performed according to ASTM E 685-93 guidelines.