Agilent 6890 Gas Chromatograph Low-Level Sulfur Analysis in Beer

This document describes the analysis and detection methods of low-level sulfur compounds in beer. The Tekmar AQUATek 70 and 3100 Sample Concentrator, along with the Agilent 6890 Gas Chromatograph and Agilent Model 355 Sulfur Chemiluminescence Detector, enable automated quantitative analysis of low-level sulfur compounds in beer. The document also compares two types of gas chromatograph columns.


Agilent 6890 Gas Chromatograph Low-Level Sulfur Analysis in Beer - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
90 KB
Published
05 July, 2012
Updated
29 September, 2023
MD5 checksum
8726EC814B407B99C9DB1AEB7F39F779

About this manual

This document details a method for the quantitative analysis of low-level volatile sulfur compounds in beer. The study utilizes an Agilent 6890 Gas Chromatograph equipped with an Agilent Model 355 Sulfur Chemiluminescence Detector (SCD) and a Tekmar AQUATek 70 and 3100 Sample Concentrator. The sample pathway is Silcosteel to prevent adsorption or decomposition of analytes.

The document provides specific operating parameters for the Tekmar 3100, AQUATek 70, and the SCD. It compares results obtained using two different gas chromatograph columns: the Agilent DB-1 and the Agilent GS-GasPro. The study reports on detection limits, repeatability, and recovery rates for various sulfur compounds found in beer.

What's inside

  • Introduction
  • Technical Overview
  • Experimental
  • Table 1. Tekmar 3100 Parameters
  • Table 2. AQUATex 70 Parameters
  • Table 3. SCD Parameters
  • Table 4. GC Parameters for the DB-1 Column
  • Table 5. GC Parameters for the GS-GasPro Column
  • Results and Discussion
  • Conclusions

Specifications

Purge time7min
Desorb temperature225°C
Burner temperature800°C
Hydrogen flow rate100 mL/ min
Air flow rate40 mU min
Column length (DB-1)30 m
Column length (GS-GasPro)60 m

Frequently asked questions

What equipment is used for the automated analysis of sulfur compounds in beer?

The analysis uses a Tekmar AQUATek 70 and 3100 Sample Concentrator coupled with an Agilent 6890 Gas Chromatograph and an Agilent Model 355 Sulfur Chemiluminescence Detector.

Which two gas chromatograph columns were compared in this study?

The study compared an Agilent GS-GasPro column and an Agilent DB-1 column.

What is the purge time for the Tekmar 3100 Sample Concentrator?

The purge time is set to 7 minutes.

Download PDF (90 KB)

More Agilent Manuals


View all Agilent manuals →