Agilent C24_Evaluation of automated sample preparation retention time locked GC-MS automated data analysis for the metabolomic study of arabidopsis species Manual

This document describes a method for automated sample preparation, retention time locked GC-MS and automated data analysis for metabolomic study


Agilent C24_Evaluation of automated sample preparation retention time locked GC-MS automated data analysis for the metabolomic study of arabidopsis species Manual - cover page
Brand
Agilent
Document type
Other
Language
English
Pages
1
File format
PDF
File size
365 KB
Published
05 July, 2012
Updated
01 October, 2023
MD5 checksum
B34CDD85B7DB3FCD717F8D57F7911910

About this manual

The document presents a scientific evaluation of an automated sample preparation method combined with retention time locked Gas Chromatography-Mass Spectrometry (GC-MS) analysis. The study focuses on the metabolomic profiling of *Arabidopsis thaliana* to assess the reproducibility and stability of derivatisation processes, specifically oximation and silylation.

The experimental setup utilized an Agilent 7693A automatic Liquid Sampler for sample preparation upstream of an Agilent 7890A GC & 5975 MSD. The study analyzed plant tissue samples from two genotypes (Wassilewskija and Columbia) at different ages, extracting metabolites using chloroform/methanol/water. Results demonstrated that automated preparation significantly improved reproducibility compared to manual methods.

What's inside

  • Introduction
  • Experimental
  • Automated Sample Preparation Sequence
  • Analytical parameters
  • Results
  • Data Analysis & Feature Detection
  • Conclusions

Frequently asked questions

What is the purpose of the automated sample preparation described in this document?

The process is used for the metabolomic study of Arabidopsis species to enhance repeatability and keep the time between sample preparation and analysis short and constant.

What are the mixing conditions for the sample preparation sequence?

The sample is mixed at 2000 rpm for 2 times, with each mix lasting 30 seconds.

What are the heating conditions for the sample preparation sequence?

The vial is heated at 30 degrees C for 3600 seconds in the first step and at 37 degrees C for 1800 seconds in the second step.