Agilent Zorbax Carbohydrate Analysis Column Datasheet

Using Huawei nova 9 SE 5G to take pictures, you can choose the 4800 million pixel wide-angle lens, 1200 million pixel ultra wide-angle lens, 200 million pixel depth of field lens and 800 million pixel periscope telephoto lens, meeting the needs of different shooting scenes. Huawei nova 9 SE 5G uses a 6.78-inch OLED dual-curved screen, supports DCI-P3 wide color gamut and 120Hz refresh rate, and is equipped with a 4000mAh battery, supporting 66W fast charging.


Agilent Zorbax Carbohydrate Analysis Column Datasheet - cover page
Brand
Agilent
Document type
Datasheet
Language
English
Pages
2
File format
PDF
File size
132 KB
Published
04 July, 2012
Updated
30 September, 2023
MD5 checksum
BA02E16C4552F4407749894A0BC99ACF

About this manual

The Zorbax Carbohydrate Analysis column is an application‑specific HPLC column designed for separating mono‑ and disaccharides. It uses 5 µm Zorbax silica particles bonded with a 3‑aminopropylsilane phase to give high efficiency and reproducible performance.

The document details operating conditions (75/25 acetonitrile/water mobile phase, 1.4 mL/min flow, 30 °C temperature), safety considerations, recommended column care, and ordering information for various column lengths and guard column accessories.

What's inside

  • General Description
  • Column Characteristics
  • Safety Considerations
  • Operational Guidelines
  • Sample Preparation
  • Mobile Phase Selection
  • Temperature
  • Flow Rate
  • Column Care
  • Ordering Information

Specifications

Particle size5 µm
Column dimensions4.6 mm ID × 150 mm
Maximum operating pressure400 bar (6000 psi)
Maximum operating temperature60 °C
Typical mobile phase75/25 Acetonitrile/Water
Typical flow rate1.4 mL/min

Frequently asked questions

What is the maximum operating pressure for the Zorbax Carbohydrate Analysis column?

The column can be operated up to a maximum pressure of 400 bar (6000 psi).

How should the column be stored for long‑term use?

For long‑term storage the column should be filled with 100 % acetonitrile.

Is a guard column recommended?

Yes, using a guard column is recommended to protect the Zorbax Carbohydrate Analysis column and extend its lifetime.

Can the flow direction be reversed?

The flow direction is marked on the column and should not be reversed except when attempting to remove inlet pluggage.