Agilent Bio-Monolith DEAE Column Data Sheet

Bio-Monolith DEAE columns are high-performance monolith columns that offer all the advantages of a specially designed, continuous short polymeric bed. Their inherent features enable highly reproducible separations of macro bio-molecules, large proteins, DNA, and viral particles, at extremely high speeds. These columns are prepacked in dedicated stainless steel housings and allow user-friendly (and quick) connections to HPLC equipment.


Agilent Bio-Monolith DEAE Column Data Sheet - cover page
Brand
Agilent
Document type
Datasheet
Language
English
Pages
2
File format
PDF
File size
37 KB
Published
04 July, 2012
Updated
30 September, 2023
MD5 checksum
BC6F1071F8E4928A82D6BCC51E8FBDFE

About this manual

The document describes the Agilent Bio-Monolith DEAE Column, catalog number 5069-3636. It is a weak anion exchanger with a poly(glycidyl methacrylate-co-ethylene dimethacrylate) monolithic matrix designed for high-speed separation of macro bio-molecules, large proteins, DNA, and viral particles.

What's inside

  • About Bio-Monolith DEAE Columns
  • Basic Characteristics
  • Taking Care of the Column
  • Regeneration
  • Cleaning in Place (CIP)
  • Long-Term Storage

Specifications

Column chemistryWeak anion exchanger; diethylamino
MatrixPoly(glycidyl methacrylate-co-ethylene dimethacrylate) highly porous monolith
Ligand density2.3 + 0.2 mmol/g dry support
Matrix dimensions (Diameter)5.2 mm
Bed volume (CV)0.10 mL
Dynamic binding capacity25 mg BSA/mL wet support
Maximum loading capacity1.2 to 14 mg
Recommended working flow rates0.2 to 2 mL/min
Maximum allowed pressure150 bar (15 MPa, 2200 psi)

Frequently asked questions

What is the recommended working temperature range?

The column should be used between 4 and 40 °C.

What is the recommended pH range for working?

The recommended working pH range is 2 to 13.

How should the column be regenerated?

Wash with at least 2 mL of a buffer containing 1M NaCl, then recondition with starting mobile phase.