Agilent LC Autosamplers and Injections Troubleshooting Video Notes

This video introduces the use and precautions of autosamplers and injectors for the Agilent 1260 and 1290 LC system.


Agilent LC Autosamplers and Injections Troubleshooting Video Notes - cover page
Brand
Agilent
Document type
User Manual
Language
English
Pages
4
File format
PDF
File size
349 KB
Published
04 July, 2012
Updated
30 September, 2023
MD5 checksum
8AB6992CF778F8B423F447D0C6B66B19

About this manual

This document provides troubleshooting guidance for Agilent Liquid Chromatography (LC) autosamplers and injection methods. It features insights from Applications Chemist Mike Woodman and Senior Applications Scientist Bill Long, focusing on common operational issues.

What's inside

  • How Autosamplers Work
  • Peaks Too Small
  • Peaks Too Large
  • Poor Reproducibility
  • Injection Volume and Peak Quality
  • Non-Linear Calibration Curves
  • Extra Peaks
  • Changing Peak Shape
  • Troubleshooting With An Autosampler Bypass
  • Summary

Frequently asked questions

What are the two main types of autosampler design?

The two main types of autosampler design are fixed loop and variable volume.

Which Agilent instruments have variable volume injectors?

Most newer instruments, including the Agilent 1260 and Agilent 1290, have variable volume injectors.

How can you validate autosampler accuracy?

To validate autosampler accuracy, install a tared vial and inject a volume that is conveniently verified gravimetrically.

What causes extra peaks in the chromatogram?

Extra peaks may be caused by contamination in the needle port or needle seat from other samples, or from the mobile phase.

Download PDF (349 KB)