Agilent PolarGel Organic GPC Columns Data Sheet

Introduces the flow rate range, recommended flow rate, sample preparation and injection information of PolarGel organic GPC columns


Agilent PolarGel Organic GPC Columns Data Sheet - cover page
Brand
Agilent
Document type
Datasheet
Language
English
Pages
4
File format
PDF
File size
274 KB
Published
04 July, 2012
Updated
30 September, 2023
MD5 checksum
3070CC0FBB216E28F24DA08DF92B0AEB

About this manual

This document provides technical information for the Agilent PolarGel Organic GPC Columns. It covers recommended installation practices, including tubing specifications and connection compatibility to ensure optimal system performance.

The guide details operational parameters such as flow rates and sample preparation requirements. It also includes specific procedures for transferring columns between different eluents and a step-by-step maintenance section for replacing frits in case of blockage.

What's inside

  • Installation
  • Eluent Flow Rate
  • Sample Preparation
  • Eluents
  • Column Testing
  • Storage
  • Maintenance

Specifications

Typical operating pressure in water450 psi (30 bar)
Maximum operating pressure2000 psi (140 bar)
Maximum operating temperature80 °C
Efficiency (H,0 at 20°C)> 35,000 plates/m

Frequently asked questions

What is the recommended flow rate for PolarGel columns using 7.6 mm id?

1.0 mL/min is an optimum flow rate for most separations.

Can the column be transferred to other eluents?

Yes, provided care is taken regarding miscibility and viscosity during transfer.

What is the warranty period for PolarGel columns?

The columns are covered by a 90-day warranty following delivery.