The Agilent 400 MHz NMR magnet, with its Premium Shielded design, is the result of continuous refinement of the SUPERSCREENED technology and delivers valuable benefits for high resolution NMR Systems.
The document describes the Agilent 400 MHz NMR Magnet featuring a Premium Shielded design based on SUPERSCREENED technology. Key benefits include significant reduction in fringe field, minimized laboratory space requirements, and outstanding homogeneity.
Technical specifications indicate an NMR frequency of 400 MHz (9.4 Tesla), with a nominal room-temperature bore of 54 mm. The system includes superconducting shim coils and features excellent fringe field characteristics to facilitate ease of siting and improve operational safety.
| NMR frequency | 400 MHz |
|---|---|
| Nominal room-temperature bore | 54 mm |
| Axial 5 G height above floor | 100 cm |
| Minimum ceiling height | 294 cm |
| Liquid helium hold time | 270 days |
The magnet has a nominal room-temperature bore of 54 mm.