Agilent 5973N Inert MSD Application Note: Using Higher Ion Source Temperatures

This article introduces the new features of Agilent Technologies, Inc.'s 5973N inert mass spectrometer detector and ChemStation software, which allows for operation of the ion source at higher temperatures, providing improvements in analysis. It also provides advice and examples on how to explore the utility of ion source temperature.


Agilent 5973N Inert MSD Application Note: Using Higher Ion Source Temperatures - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
943 KB
Published
02 July, 2012
Updated
29 September, 2023
Model
5973N
MD5 checksum
6F9FBB8C4FFE56BC813C4EC6C9F1E7FF

About this manual

This document discusses the capability of the Agilent 5973N Inert MSD and ChemStation software to operate the ion source at temperatures up to 300 °C, compared to the standard 230 °C. It details how higher temperatures affect tuning results, background noise, and signal-to-noise ratios for specific analytes like polychlorinated biphenyls (PCBs) and coronene. The note highlights that while higher temperatures can improve response for late-eluting compounds, they also increase background noise, which may lower the signal-to-noise ratio in full-scan mode but can be beneficial in selected-ion-monitoring (SIM) mode for certain ions.

What's inside

  • Introduction
  • Tuning
  • Implications for Analytical Applications
  • Source Bakeout
  • ChemStation Macro for Baking
  • Pumpdown Sequence

Specifications

Maximum Ion Source Temperature300 °C
Default Ion Source Temperature230 °C
Maximum Quadrupole Temperature200 °C

Frequently asked questions

What is the maximum ion source temperature for the 5973N Inert MSD?

The ion source temperature can be set to a maximum of 300 °C.

Does operating at higher source temperatures improve the signal-to-noise ratio for all compounds?

No, while signal increases, background noise also increases, which can result in a lower signal-to-noise ratio for full-scan acquisitions of certain compounds.

Download PDF (943 KB)