Agilent Liberty Series ICP-OES Application Note for Potable Water Analysis

Analysis of Potable Water for Trace Elements by ICP-OES


Agilent Liberty Series ICP-OES Application Note for Potable Water Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
225 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
22A075BFF6E99BBB391944EA7EC53614

About this manual

The document details a method for determining major, minor, and trace elements in water samples using Inductively Coupled Plasma-Optical Emission Spectroscopy (ICP-OES). To ensure accurate results near detection limits, the study employed a ten-fold pre-concentration of samples via evaporation. The analysis was validated against USEPA Pollution Quality Control water samples.

What's inside

  • Introduction
  • Instrumental Parameters
  • Sample Preparation
  • Standard Preparation
  • Analytical Line Selection
  • Detection Limits
  • Results

Specifications

RF Generator Frequency40.68 MHz
Power1.2 kW
Plasma gas flow12.0 L/min
Sample uptake rate1.4 mL/min

Frequently asked questions

Which Agilent instruments are used in this study?

Measurements were performed on either the Agilent Liberty 100 or the Liberty 200 Inductively Coupled Plasma Atomic Emission Spectrometer.

What pre-treatment was applied to the water samples?

A ten-fold pre-concentration by evaporation was applied to avoid routine analysis close to the detection limit.

What is the detection limit for Aluminum (Al) at 167.081 nm?

The detection limit for Al at 167.081 nm is 15 ng/mL.