Agilent 1100 Series HPLC Technical Note for Operation of Two Autosamplers

This note describes the operation of two Agilent 1100 autosamplers in a 1100 system. In order to achieve a high sample throughput both autosamplers are used. While one of the autosamplers is injecting, the other one is in the by-pass position and vice versa.


Agilent 1100 Series HPLC Technical Note for Operation of Two Autosamplers - cover page
Brand
Agilent
Document type
Other
Language
English
Pages
4
File format
PDF
File size
241 KB
Published
08 June, 2012
Updated
28 September, 2023
MD5 checksum
0BF950030BC89005FAB39630F680C8D8

About this manual

The document explains how to operate two Agilent 1100 autosamplers simultaneously within one HPLC system. By using specific macros (MALS1.MAC and MALS2.MAC), the system ensures that while one autosampler is injecting, the other remains in a bypass position to minimize dead volume.

What's inside

  • Function description of the macros
  • Listing of macro MALS1.MAC
  • Listing of macro MALS2.MAC
  • Automation example for 200 injections
  • Run-time checklist entries