Agilent Using the Test Chips Manual

This document introduces the hardware diagnostic tool for the Agilent 2100 bioanalyzer system


Agilent Using the Test Chips Manual - cover page
Brand
Agilent
Document type
User Manual
Language
English
Pages
4
File format
PDF
File size
58 KB
Published
08 June, 2012
Updated
30 September, 2023
MD5 checksum
4919D098C3CAD8BCEF307858252AA6B6

About this manual

This manual describes a software-supported hardware diagnostic tool included in all Agilent 2100 bioanalyzer systems. The tool allows users to check individual hardware components and the instrument-LabChip interface without assistance from service personnel. Results are reported as 'passed' or 'failed', with failed tests indicating faulty components requiring service attention.

What's inside

  • General
  • How to Start the Hardware Diagnostic Tool
  • Using the Autofocus Chip - Autofocus Test
  • Chip Handling and Maintenance (Autofocus)
  • Using the Electrode / Diode Chip - Electrode / Diode Test
  • Chip Handling and Maintenance (Electrode/Diode)
  • Using the Leak Current Chip - Leak Current Test
  • Diagnostic Test (Leak Current)

Specifications

Autofocus Chip Replacement IntervalTwo years from first usage

Frequently asked questions

Do I need service personnel to run the hardware diagnostic tool?

No help from Agilent service personnel is necessary to run the hardware diagnostic tool.

What indicates a faulty hardware component during testing?

A 'failed' hardware test result indicates a faulty hardware component and requires contacting Agilent service personnel.

How often should the Autofocus chip be replaced?

The Autofocus chip must be replaced by a new one after two years from first usage.