Agilent Fast LC Column User Manual for Pursuit XRs 2.8 and Pursuit UPS

This manual provides guidelines for optimizing the performance of the Pursuit™XRs Ultra 2.8 and Pursuit™UPS2.4 HPLC columns for high speed separations. The manual provides information on reducing system dead-volume and other band broadening contributions to achieve maximum separation efficiency with these low-volume columns.


Agilent Fast LC Column User Manual for Pursuit XRs 2.8 and Pursuit UPS - cover page
Brand
Agilent
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
78 KB
Published
07 June, 2012
Updated
30 September, 2023
MD5 checksum
1B3DDBB035E960F9F93279694D00B172

About this manual

The document outlines specific requirements for the HPLC system to achieve high efficiency with small particle size columns, such as minimizing dead volume in connecting tubing and detector flow cells. It details best practices for injector design, sample solvent preparation, and injection volumes to prevent band broadening.

Technical specifications are provided for two column series: the Pursuit UPS 24 supports a maximum pressure of up to 400 bar (5800 psi), while the Pursuit XRs 2.8 supports up to 1034 bar (15,000 psi).

What's inside

  • Guidelines for Optimizing Performance
  • Connecting Tubing
  • Detector Requirements
  • Injector Design
  • Sample Solvent Preparation
  • Injection Volume
  • Maximum Pressure Specifications

Specifications

Maximum Pressure (Pursuit UPS 24)up to 400 bar/5800 psi
Maximum Pressure (Pursuit XRs 2.8)up to 1034 bar/15,000 psi

Frequently asked questions

What tubing internal diameter is recommended for connecting the column?

Use narrow internal diameters of at most 0.010-inch (0.25 mm) to minimize dead volume.

What is the recommended detector flow cell volume?

The detector flow cells must have a small internal volume, preferably less than 2 uL.