Agilent Hi-Plex Ligand Exchange Columns Installation and Maintenance Guide

The document is the Hi-Plex column installation and maintenance instructions, which introduces the connection and installation, transportation liquid, column conditioning and other contents of the Hi-Plex column.


Agilent Hi-Plex Ligand Exchange Columns Installation and Maintenance Guide - cover page
Brand
Agilent
Document type
User Manual
Language
English
Pages
4
File format
PDF
File size
308 KB
Published
07 June, 2012
Updated
30 September, 2023
MD5 checksum
937EDB998863A43B72658D26378694E7

About this manual

This document provides operational guidance for the installation and connection of Hi-Plex ligand exchange columns. It details proper tightening procedures for end fittings and specifies requirements for shipping eluent handling to prevent column drying.

The guide covers mobile phase preparation, sample filtration, and routine maintenance protocols including efficiency testing and column clean-up procedures. It also outlines system shutdown steps and long-term storage conditions.

What's inside

  • Installation
  • Column Connections
  • Shipping Eluent
  • Column Conditioning
  • Mobile Phases
  • Sample Preparation
  • Maintenance
  • Column Clean-Up
  • System Shutdown

Frequently asked questions

How should Hi-Plex column end fittings be tightened?

Nuts should be tightened finger tight and then 1/4 turn using a wrench.

What is the recommended flow rate for column conditioning?

The mobile phase should be run at 0.1 mL/min during initial conditioning.

Can Hi-Plex columns be stored in a refrigerator?

Yes, they may be stored in a refrigerator at 4 °C but must not freeze.