Agilent Chip Priming Station 5065-4401 User Manual

The Chip Priming Station is an accessory for the Agilent 2100 Bioanalyzer LabChip® Kits. It can be used to fill all channels of the chip with gel or a gel/dye mixture.


Agilent Chip Priming Station 5065-4401 User Manual - cover page
Brand
Agilent
Document type
User Manual
Language
English
Pages
8
File format
PDF
File size
302 KB
Published
06 June, 2012
Updated
01 October, 2023
MD5 checksum
B54384FAD2A99AF7218B782BD34DD672

About this manual

The Agilent Chip Priming Station (part number 5065-4401) is designed for use with the Agilent 2100 Bioanalyzer LabChip® Kits. It facilitates the application of gel or dye mixtures into chip channels by creating a tight seal between the syringe and the chip well, preventing air leaks and bubble formation.

What's inside

  • Assembling the Device
  • Adjusting Base Plate
  • Placing Chip
  • Seal Test
  • Using the Station
  • Maintenance
  • Replacing Syringe Adapter
  • Gasket Replacement

Frequently asked questions

How often should maintenance be performed?

Maintenance steps, such as replacing the syringe adapter and gasket, should be performed on a regular 6 months basis.

What indicates that the seal test has failed?

A failed seal test is indicated if the plunger does not move up to the 0.3ml mark within one second.