Rosemount 370XA Error 9.1.15: Stream (1, 2, or 3) RF Deviation

Rosemount 370XA gas chromatograph · taken from the manufacturer’s manual, pages cited

Alarm code9.1.15

Stream (1, 2, or 3) RF Deviation alarm indicates streams are incorrectly configured for calibration. The manual's first action is: “Start MON2020 and select Streams on the Application menu.”

Model
Rosemount 370XA
Display shows
Stream (1, 2, or 3) RF Deviation (9.1.15)
Type
Alarm code
Source
Manual p. 159

What 9.1.15 means

Stream (1, 2, or 3) RF Deviation p. 159

Common causes

  • Streams 1 through 3 should only be set to Analysis or Unused. This alarm may occur if the streams are incorrectly configured for calibration. p. 159

What to do

  1. Start MON2020 and select Streams on the Application menu. p. 159
  2. Set the stream's Usage to Analysis or Unused. p. 159

Before you service the product, read the safety instructions in the manual p. 3.

If 9.1.15 is still not resolved

  • Start MON2020 and select Streams on the Application menu. p. 159
    “Start MON2020 and select Streams on the Application menu.”
  • Set the stream's Usage to Analysis or Unused. p. 159
    “Set the stream's Usage to Analysis or Unused.”

This page summarises what the manufacturer’s manual says about 9.1.15; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Rosemount. Found a mistake? Report an error.