Rosemount 370XA Error 9.1.14: Stream Validation Failure

Rosemount 370XA gas chromatograph · taken from the manufacturer’s manual, pages cited

Alarm code9.1.14

The analysis of the validation gas for the associated stream is outside the allowable percent deviation, as defined in the Validation Data table. The manual's first action is: “Confirm that the validation gas concentration is entered correctly in the Validation Data table.”

Model
Rosemount 370XA
Display shows
Stream Validation Failure (9.1.14)
Type
Alarm code
Source
Manual p. 159

What 9.1.14 means

Stream Validation Failure p. 159

Common causes

  • The analysis of the validation gas for the associated stream is outside the allowable percent deviation, as defined in the Validation Data table. p. 159

What to do

  1. Confirm that the validation gas concentration is entered correctly in the Validation Data table. p. 159
  2. If the concentrations are entered correctly, check the validation run analysis results and chromatogram for analysis errors. p. 159

Before you service the product, read the safety instructions in the manual p. 3.

If 9.1.14 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Rosemount service.

This page summarises what the manufacturer’s manual says about 9.1.14; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Rosemount. Found a mistake? Report an error.