Rosemount 370XA Error 9.1.13: Analog (Input or Output) (High or Low) Signal

Rosemount 370XA gas chromatograph · taken from the manufacturer’s manual, pages cited

Alarm code9.1.13

Analog input or output signal is outside the allowed range for the analog output. The manual's first action is: “If the value is valid, re-range the analog output and the associated device that is receiving the signal so that the value of the variable is within the range.”

Model
Rosemount 370XA
Display shows
Analog (Input or Output) (High or Low) Signal (9.1.13)
Type
Alarm code
Source
Manual p. 159

What 9.1.13 means

Analog (Input or Output) (High or Low) Signal p. 159

Common causes

  • The value of the variable assigned to the analog output is outside the range of the analog output. p. 159

What to do

  1. If the value is valid, re-range the analog output and the associated device that is receiving the signal so that the value of the variable is within the range. p. 159

Before you service the product, read the safety instructions in the manual p. 3.

If 9.1.13 is still not resolved

  • Contact service or dealer for assistance. p. 159
    “If the value is valid, re-range the analog output and the associated device that is receiving the signal so that the value of the variable is within the range.”

This page summarises what the manufacturer’s manual says about 9.1.13; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Rosemount. Found a mistake? Report an error.