The electrosurgical generator has fallen below its minimum operating temperature. The manual's first action is: “Switch off the electrosurgical generator and wait until it has reached the specified operating temperatu…
The electrosurgical generator has malfunctioned. The manual's first action is: “If the problem persists, contact Olympus.”
The electrosurgical generator has malfunctioned. The manual's first action is: “If the problem persists, contact Olympus.”
The HF instrument or connection cable has malfunctioned. The manual's first action is: “Replace the HF instrument and / or the connection cable.”
The HF instrument or connection cable has malfunctioned. The manual's first action is: “Replace the HF instrument and / or the connection cable.”
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket. The manual's first action is: “Check the connection of the Olympus HF instrument to the UNIVERSAL socket.”
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket. The manual's first action is: “Check the connection of the Olympus HF instrument to the UNIVERSAL socket.”
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket. The manual's first action is: “Check the connection of the Olympus HF instrument to the UNIVERSAL socket.”
The Olympus HF instrument is not properly connected to the UNIVERSAL socket. The manual's first action is: “Check the connection of the Olympus HF instrument to the UNIVERSAL socket.”
The power level for the chosen mode is set to zero. The manual's first action is: “Increase the power level via the "Set screen" (refer to chapter 5.4, "Output setting").”
The device indicates no mode has been selected while activating. The manual's first action is: “Select a mode via the "Mode screen" (refer to chapter 5.4, "Output setting").”
The double pedal footswitch has not been assigned to the corresponding output socket. The manual's first action is: “Assign the double pedal footswitch via the "Assign Footswitch screen" (refer to chapter 6.4, "Assign…
The single pedal footswitch has not been assigned to the corresponding output socket. The manual's first action is: “Assign the single pedal footswitch via the "Assign Footswitch screen" (refer to chapter 6.4, "Assign…
The maximum time limit for the application has been exceeded. The manual's first action is: “Release the footswitch or handswitch and activate again by repressing the footswitch or handswitch.”
The screen is touched while switching on or there's a malfunction of the electrosurgical generator. The manual's first action is: “Release the finger from the screen.”