Communication error E490 indicates improper cable connection or cable damage. The manual's first action is: “Check the connection of the communication cables to the electrosurgical generator and / or to the peripheral…
Communication error E411 indicates improper cable connection or cable damage. The manual's first action is: “Check the connection of the communication cables to the electrosurgical generator and / or to the peripheral…
Communication error E409 indicates improper cable connection or cable damage. The manual's first action is: “Check the connection of the communication cables to the electrosurgical generator and / or to the peripheral…
Communication error indicates improper cable connection or cable damage. The manual's first action is: “Check the connection of the communication cables to the electrosurgical generator and / or to the peripheral equi…
E390 indicates a communication error between the device and its peripherals. The manual's first action is: “Check the connection of the communication cables to the electrosurgical generator and / or to the peripheral …
The electrosurgical generator has detected a low battery condition. The manual's first action is: “If the problem persists, contact Olympus.”
Excessive HF leakage current has exceeded safety limits. The manual's first action is: “Check if the electrosurgical generator or the patient is grounded unintentionally.”
The high frequency leakage current has exceeded the limit for monopolar or bipolar application. The manual's first action is: “Check if the electrosurgical generator or the patient is grounded unintentionally.”
The maximum temperature has been exceeded or there is a malfunction of the electrosurgical generator. The manual's first action is: “Switch off the electrosurgical generator and wait until it has cooled down or reache…
The maximum temperature has been exceeded or there is a malfunction of the electrosurgical generator. The manual's first action is: “Switch off the electrosurgical generator and wait until it has cooled down or reache…
The maximum temperature has been exceeded or there is a malfunction of the electrosurgical generator. The manual's first action is: “Switch off the electrosurgical generator and wait until it has cooled down or reache…
The maximum temperature has been exceeded. The manual's first action is: “Switch off the electrosurgical generator and wait until it has cooled down or reached the specified operating temperature (refer to "Transporta…
The electrosurgical generator has fallen below the minimum operating temperature or malfunctioned. The manual's first action is: “Switch off the electrosurgical generator and wait until it has reached the specified op…
The electrosurgical generator has fallen below the minimum operating temperature or malfunctioned. The manual's first action is: “Switch off the electrosurgical generator and wait until it has reached the specified op…
The electrosurgical generator has fallen below the minimum operating temperature or malfunctioned. The manual's first action is: “Switch off the electrosurgical generator and wait until it has reached the specified op…